Anaerobic Chamber Service & Repair

Anaerobic chambers and workstations depend on tight seals, the right gas mix and accurate oxygen monitoring. Nexus Biomedical Services repairs and maintains anaerobic systems for clinical microbiology and research labs across New Jersey, Pennsylvania, New York, Texas and Delaware.

Anaerobic systems we service

Anaerobic chambers and workstations with airlocks, glove ports or sleeves, catalysts and oxygen sensors, from the major manufacturers. Share the model and serial number and we will confirm support before booking a visit.

Common faults we fix

  • Oxygen levels rising or slow to drop
  • Leaking glove ports, sleeves, seals or doors
  • Airlock cycling and vacuum faults
  • Gas supply, regulator and valve problems
  • Oxygen sensor and display errors
  • Heating, humidity and condensation problems

Preventive maintenance

A planned visit typically covers leak checks of seals, gloves and doors, airlock and vacuum checks, gas supply and regulator inspection, catalyst and sensor checks, temperature checks and electrical safety testing. Most labs schedule preventive maintenance every 6 or 12 months.

Atmosphere and safety checks

After service we confirm the chamber holds its atmosphere, check the oxygen sensor reading and temperature, and record the results along with electrical safety testing.

What to send us

Manufacturer, model, serial number and asset ID; oxygen readings; gas mix used; and when the problem started. Then request service or call (848) 405-6968.

Frequently asked questions

Leaking gloves, seals or the airlock are common causes, along with an exhausted catalyst. We leak-check each.

Yes, at every visit.

Every 6 to 12 months for most labs.