Autoclave Service & Repair

Laboratory autoclaves sterilize media, glassware and waste, and a failed cycle can stop work or put staff at risk. Nexus Biomedical Services repairs and maintains benchtop and floor-standing laboratory autoclaves for clinical, microbiology and research labs across New Jersey, Pennsylvania, New York, Texas and Delaware.

Autoclaves we service

Benchtop and floor-standing laboratory steam autoclaves from the major manufacturers. Share the model and serial number and we will confirm support before booking a visit.

Common faults we fix

  • Cycles aborting or failing to reach temperature or pressure
  • Door gasket leaks, latch and interlock faults
  • Steam generator, heating element and water supply problems
  • Drain, exhaust and condensate faults
  • Temperature and pressure sensor errors
  • Printer, display and cycle record faults

Preventive maintenance

A planned visit typically covers door gasket and interlock checks, chamber and drain cleaning, steam generator and heating checks, safety valve inspection, temperature and pressure sensor checks against calibrated references, a test cycle and electrical safety testing. Most labs schedule preventive maintenance every 6 or 12 months.

Temperature, pressure and safety verification

After service we compare chamber temperature and pressure with calibrated references during a test cycle and record the results. Biological and chemical indicator testing stays with your lab’s sterilization program.

What to send us

Manufacturer, model, serial number and asset ID; the cycle and step where it failed; error codes; and recent indicator results. Then request service or call (848) 405-6968.

Frequently asked questions

Common causes include door seals, drain blockages, heating problems and sensor faults. We check each before replacing parts.

Yes. Door interlocks are checked at every visit.

Every 6 to 12 months for most labs.